Dark kitchen during a storm power outage, with a portable power station keeping the fridge and a lamp running

A CPAP night costs anywhere from about 50 Wh to more than 500 Wh. The humidifier and heated tube decide the size of power station you need, not the machine.

Shopping for a station? Start with the CPAP power station buying guide: picks for outages and camping, UPS switchover times, ResMed DC converter limits, and nights that include the inverter's own draw.

ResMed's power table measures its machines at about 6.3–9 W with no humidifier, 19–34 W with a humidifier, and 34–66 W with a humidifier and heated tube (ResMed). Those bands are why a night runs from about 50 Wh to more than 500 Wh. A car battery is not a power station: can a car battery run a CPAP separates the DC example from the inverter case.

Medical safety notice: Will It Run? gives general power estimates only, not medical advice. Power draw varies by model, settings, pressure, humidifier and altitude. Before relying on any battery, power station, inverter or generator for a medical device, confirm approved power sources with the device manufacturer and your doctor or equipment supplier, and test your setup before an emergency. Never change prescribed settings to extend battery life. If your therapy is life-sustaining, keep the backup your supplier provides, register with your utility's medical-needs program, and have an evacuation plan. Run generators outdoors only, at least 20 ft from windows and doors, because of carbon monoxide risk. In an emergency, call 911.

Every machine and setting

Machine and settingAverage draw8-hour nightNights on a 1,024 Wh station, ACOn DC, best caseSource
ResMed AirMini travel CPAP6.3 W50 Wh5.117.3ResMed
ResMed AirSense 11 AutoSet, no humidifier7 W56 Wh4.915.5ResMed
ResMed AirSense 10 AutoSet, no humidifier7 W56 Wh4.915.5ResMed
ResMed AirSense 11 with HumidAir 11 humidifier, no heated tube19 W152 Wh3.25.7ResMed
ResMed AirSense 10 AutoSet with HumidAir humidifier22 W176 Wh2.9—ResMed
ResMed AirSense 10 with HumidAir and ClimateLineAir heated tube34 W272 Wh2.2—ResMed
ResMed AirSense 11 with HumidAir 11 and ClimateLineAir 1155 W440 Wh1.62.0ResMed
ResMed AirSense 11, typical power from the user guide56.1 W449 Wh1.51.9ResMed
Philips DreamStation, worst case at the 80 W power supply rating80 W640 Wh1.1—Philips

AC nights = 1,024 Wh × 0.85 usable ÷ ((watts + about 15 W inverter draw) × 8 hr). DC nights = 1,024 Wh × 0.85 ÷ (watts × 8 hr), a best case with the maker's DC converter (ResMed Air11, Air10 or AirMini) in a 12 V car port; — means DC isn't modeled for that setting (ResMed's AirSense 10 guide says humidity off on an external battery, and Philips publishes no DC figure). ResMed figures are lab measurements from ResMed's published power table; the Philips row is the 80 W power-supply ceiling, a worst case. The 15 W is the inverter's own draw: BLUETTI counts about 15 W of self-consumption in its AC180 runtime formula, and an owner test measured about 17.5 W on a DELTA 3 Plus; larger stations can idle higher. Method, DC converter limits and picks: CPAP power station guide. The RIVER 2, RIVER 3 Plus and SOLIX C300 DC figures use their 12 V car outputs (EcoFlow RIVER 2 manual: 100 W max; RIVER 3 Plus specs: 126 W max; Anker C300 guide: 12 V 10 A). The DELTA Pro 3 has 12 V Anderson and DC5521 outputs but no car socket, so it stays AC-only here.

BiPAP (ResMed AirCurve)

Machine and settingAverage draw8-hour nightNights on a 1,024 Wh station, ACOn DC, best caseSource
ResMed AirCurve 11 VAuto BiPAP, no humidifier8 W64 Wh4.713.6ResMed
ResMed AirCurve 10 VAuto BiPAP, no humidifier9 W72 Wh4.512.1ResMed
ResMed AirCurve 10 VAuto BiPAP with HumidAir humidifier23 W184 Wh2.9—ResMed
ResMed AirCurve 11 VAuto BiPAP with HumidAir humidifier34 W272 Wh2.23.2ResMed
ResMed AirCurve 10 VAuto BiPAP with HumidAir and heated tube41 W328 Wh1.9—ResMed
ResMed AirCurve 11 VAuto BiPAP with HumidAir and heated tube66 W528 Wh1.31.6ResMed

AC nights = 1,024 Wh × 0.85 usable ÷ ((watts + about 15 W inverter draw) × 8 hr). DC nights = 1,024 Wh × 0.85 ÷ (watts × 8 hr), a best case with the maker's DC converter (ResMed Air11, Air10 or AirMini) in a 12 V car port; — means DC isn't modeled for that setting (ResMed's AirSense 10 guide says humidity off on an external battery, and Philips publishes no DC figure). ResMed figures are lab measurements from ResMed's published power table; the Philips row is the 80 W power-supply ceiling, a worst case. The 15 W is the inverter's own draw: BLUETTI counts about 15 W of self-consumption in its AC180 runtime formula, and an owner test measured about 17.5 W on a DELTA 3 Plus; larger stations can idle higher. Method, DC converter limits and picks: CPAP power station guide.

Nights on each station (AirSense 11)

StationCapacityNo humidifier (AC)Humidifier (AC)Humidifier + heated tube (AC)No humidifier, DC best casePrice
EcoFlow RIVER 2256 Wh1.20.80.43.9Amazon
EcoFlow RIVER 3 Plus286 Wh1.40.90.44.3Amazon
Anker SOLIX C300288 Wh1.40.90.44.4Amazon
Goal Zero Yeti 500X497 Wh2.41.60.87.5Amazon
EcoFlow RIVER 2 Max512 Wh2.51.60.87.8Amazon
EcoFlow RIVER 2 Pro768 Wh3.72.41.211.7Amazon
BLUETTI AC70768 Wh3.72.41.211.7Amazon
Anker SOLIX C800768 Wh3.72.41.211.7Amazon
Goal Zero Yeti 1000X997 Wh4.83.11.515.1Amazon
EcoFlow DELTA 21,024 Wh4.93.21.615.5Amazon
EcoFlow DELTA 31,024 Wh4.93.21.615.5Amazon
EcoFlow DELTA 3 Plus1,024 Wh4.93.21.615.5Amazon
Anker SOLIX C10001,056 Wh5.13.31.616.0Amazon
Jackery Explorer 1000 v21,070 Wh5.23.31.616.2Amazon
BLUETTI AC1801,152 Wh5.63.61.717.5Amazon
Jackery Explorer 1000 Plus1,264.64 Wh6.14.01.919.2Amazon
Goal Zero Yeti 1500X1,516 Wh7.34.72.323.0Amazon
Jackery Explorer 2000 v22,042 Wh9.96.43.131.0Amazon
Jackery Explorer 2000 Plus2,042.8 Wh9.96.43.131.0Amazon
EcoFlow DELTA 2 Max2,048 Wh9.96.43.131.1Amazon
BLUETTI AC200L2,048 Wh9.96.43.131.1Amazon
Anker SOLIX F2000 (PowerHouse 767)2,048 Wh9.96.43.131.1Amazon
Anker SOLIX F38003,840 Wh8.76.94.358.3Amazon
EcoFlow DELTA Pro 34,096 Wh19.812.86.2—Amazon

Anker SOLIX F3800 row: nights use about 40 W of idle draw instead of 15 W, because Anker's FAQ for the F3800 Plus lists about 40 W with AC on and nothing connected (Anker). At CPAP loads that cuts its nights by roughly a third to a half; see the F3800 idle table. The F3800 DC column uses its 12 V ⎓ 10 A car charger output (Anker US user guide), which skips the inverter.

Nights of 8 hours. AC = nameplate Wh × 0.85 ÷ (average draw + about 15 W inverter draw). DC = nameplate Wh × 0.85 ÷ average draw, a best case with ResMed's Air11 DC converter in the 12 V car port; — means no car port is listed in the maker specs we checked. As an Amazon Associate we earn from qualifying purchases. The 15 W is the inverter's own draw: BLUETTI counts about 15 W of self-consumption in its AC180 runtime formula, and an owner test measured about 17.5 W on a DELTA 3 Plus; larger stations can idle higher. Method, DC converter limits and picks: CPAP power station guide.

Skipped models

Philips DreamStation 2, React Health Luna G3, Philips BiPAP models and the Transcend travel CPAP are not listed yet because their makers do not publish a power consumption figure we could verify. We would rather leave a model out than guess.

More CPAP guides

Questions

How much power does a CPAP use?

ResMed measures 6.3 to 9 W for its machines without a humidifier, 19 to 34 W with a humidifier, and 34 to 66 W with a humidifier and heated tube.

Which models are missing?

DreamStation 2, Luna G3, Philips BiPAP and Transcend, because no verifiable manufacturer power figure was found.