How many watts does a CPAP use?

Portable power station on a kitchen counter charging a phone next to a coffee maker

ResMed measures its CPAPs at 6.3 to 7 W without a humidifier, 19 to 22 W with a humidifier, and 34 to 55 W with a heated tube.

Medical safety notice: Will It Run? gives general power estimates only, not medical advice. Power draw varies by model, settings, pressure, humidifier and altitude. Before relying on any battery, power station, inverter or generator for a medical device, confirm approved power sources with the device manufacturer and your doctor or equipment supplier, and test your setup before an emergency. Never change prescribed settings to extend battery life. If your therapy is life-sustaining, keep the backup your supplier provides, register with your utility's medical-needs program, and have an evacuation plan. Run generators outdoors only, at least 20 ft from windows and doors, because of carbon monoxide risk. In an emergency, call 911.

Machine and settingAverage draw8-hour nightNights on a 1,024 Wh station, ACOn DC, best caseSource
ResMed AirMini travel CPAP6.3 W50 Wh5.117.3ResMed
ResMed AirSense 11 AutoSet, no humidifier7 W56 Wh4.915.5ResMed
ResMed AirSense 10 AutoSet, no humidifier7 W56 Wh4.915.5ResMed
ResMed AirSense 11 with HumidAir 11 humidifier, no heated tube19 W152 Wh3.25.7ResMed
ResMed AirSense 10 AutoSet with HumidAir humidifier22 W176 Wh2.9—ResMed
ResMed AirSense 10 with HumidAir and ClimateLineAir heated tube34 W272 Wh2.2—ResMed
ResMed AirSense 11 with HumidAir 11 and ClimateLineAir 1155 W440 Wh1.62.0ResMed
ResMed AirSense 11, typical power from the user guide56.1 W449 Wh1.51.9ResMed
Philips DreamStation, worst case at the 80 W power supply rating80 W640 Wh1.1—Philips

AC nights = 1,024 Wh × 0.85 usable ÷ ((watts + about 15 W inverter draw) × 8 hr). DC nights = 1,024 Wh × 0.85 ÷ (watts × 8 hr), a best case with the maker's DC converter (ResMed Air11, Air10 or AirMini) in a 12 V car port; — means DC isn't modeled for that setting (ResMed's AirSense 10 guide says humidity off on an external battery, and Philips publishes no DC figure). ResMed figures are lab measurements from ResMed's published power table; the Philips row is the 80 W power-supply ceiling, a worst case. The 15 W is the inverter's own draw: BLUETTI counts about 15 W of self-consumption in its AC180 runtime formula, and an owner test measured about 17.5 W on a DELTA 3 Plus; larger stations can idle higher. Method, DC converter limits and picks: CPAP power station guide.

The power supply rating on the brick (65 W, 80 W or 90 W) is a maximum, not the typical draw. ResMed's AirSense 11 user guide lists a 56.1 W typical and 73.2 W peak system figure, which Willrun uses as its default AirSense 11 row. To measure your own machine, a plug-in watt meter gives the real number at your settings.

BiPAP

Machine and settingAverage draw8-hour nightNights on a 1,024 Wh station, ACOn DC, best caseSource
ResMed AirCurve 11 VAuto BiPAP, no humidifier8 W64 Wh4.713.6ResMed
ResMed AirCurve 10 VAuto BiPAP, no humidifier9 W72 Wh4.512.1ResMed
ResMed AirCurve 10 VAuto BiPAP with HumidAir humidifier23 W184 Wh2.9—ResMed
ResMed AirCurve 11 VAuto BiPAP with HumidAir humidifier34 W272 Wh2.23.2ResMed
ResMed AirCurve 10 VAuto BiPAP with HumidAir and heated tube41 W328 Wh1.9—ResMed
ResMed AirCurve 11 VAuto BiPAP with HumidAir and heated tube66 W528 Wh1.31.6ResMed

AC nights = 1,024 Wh × 0.85 usable ÷ ((watts + about 15 W inverter draw) × 8 hr). DC nights = 1,024 Wh × 0.85 ÷ (watts × 8 hr), a best case with the maker's DC converter (ResMed Air11, Air10 or AirMini) in a 12 V car port; — means DC isn't modeled for that setting (ResMed's AirSense 10 guide says humidity off on an external battery, and Philips publishes no DC figure). ResMed figures are lab measurements from ResMed's published power table; the Philips row is the 80 W power-supply ceiling, a worst case. The 15 W is the inverter's own draw: BLUETTI counts about 15 W of self-consumption in its AC180 runtime formula, and an owner test measured about 17.5 W on a DELTA 3 Plus; larger stations can idle higher. Method, DC converter limits and picks: CPAP power station guide.

Kill A Watt meter on AmazonAs an Amazon Associate we earn from qualifying purchases.

More CPAP guides

Questions

How many watts does a CPAP use?

ResMed measures its CPAPs at 6.3 to 7 W without a humidifier, 19 to 22 W with a humidifier, and 34 to 55 W with a heated tube.

Is the power supply rating the real draw?

No. It is the maximum the supply can deliver. Typical draw is lower.