Answer

Can the Goal Zero Yeti 500X run ResMed AirSense 10 AutoSet, no humidifier?

Runs

Runs. Draw is 7 W, within the 300 W continuous output. Estimated runtime on AC: about 19.2 hr, counting about 15 W of inverter draw.

Medical safety notice: Will It Run? gives general power estimates only, not medical advice. Power draw varies by model, settings, pressure, humidifier and altitude. Before relying on any battery, power station, inverter or generator for a medical device, confirm approved power sources with the device manufacturer and your doctor or equipment supplier, and test your setup before an emergency. Never change prescribed settings to extend battery life. If your therapy is life-sustaining, keep the backup your supplier provides, register with your utility's medical-needs program, and have an evacuation plan. Run generators outdoors only, at least 20 ft from windows and doors, because of carbon monoxide risk. In an emergency, call 911.

Check price on AmazonManufacturer pageAs an Amazon Associate we earn from qualifying purchases.

The comparison

Goal Zero Yeti 500XResMed AirSense 10 AutoSet, no humidifier
Continuous output / running draw300 W7 W
Capacity / average draw497 Wh nameplate, 422 Wh usable7 W
Runtime on AC19.2 hr (7 W + about 15 W inverter draw)
Nights on AC (8 hr)2.4
On DC, best case2.5 days, 7.5 nights with the ResMed Air10 DC/DC converter 90W in the 12 V car port
SourcesStation sourceAppliance source

How AC runtime is figured: usable Wh (nameplate × 0.85) ÷ (CPAP watts + about 15 W). The 15 W is the inverter's own draw just to stay on: BLUETTI counts about 15 W of self-consumption in its AC180 runtime formula, and an owner test measured about 17.5 W idle on a DELTA 3 Plus. Larger stations can idle higher, so test your own setup before you rely on it. DC figures skip the inverter and are a best case. Method and DC converter limits: CPAP power station guide.

More on the Goal Zero Yeti 500X · More on this machine · CPAP hub

Questions

Can the Goal Zero Yeti 500X run ResMed AirSense 10 AutoSet, no humidifier?

Runs. Draw is 7 W, within the 300 W continuous output. Estimated runtime on AC: about 19.2 hr, counting about 15 W of inverter draw.

How was the runtime calculated?

On AC: 497 Wh × 0.85 = 422 Wh usable, divided by (7 W + about 15 W inverter draw) = 19.2 hr. On DC with the ResMed Air10 DC/DC converter 90W in the car port there is no inverter, so the best case is 422 Wh divided by 7 W = 2.5 days.

Where did the numbers come from?

Spec list: 497 Wh, Li-ion NMC, 500 cycles to 80% at 0.2C, 12.9 lb, 2× 120 V AC at 300 W and 1,200 W surge. A shelf price on the page was not recorded as a launch price. ResMed's therapy-device power table (ResMed Deutschland, Art.-Nr. 1018528/11) lists AirSense 10 AutoSet at 7 Wh during operation. Energy per hour equals average watts, so the average draw is 7 W. ResMed notes the values were measured in a lab and that a local meter can differ.